Carbon Monoxide, 2 - 1000 ppm (8610080)
Carbon Monoxide, 2 - 1000 ppm, Part # 8610080
Measuring Range:
2 to 1000 ppm
Typical Measuring Time:
approx. 5 to 300 s
Accuracy:
± 25 % (under calibration conditions)
Ambient Conditions:
Temperature:
10 to 40 °C
Temperature correction:
none
Humidity:
1 to 40 mg/L
Humidity correction:
none
Cross Sensitivity:
No influence at 10 ppm carbon monoxide by:
2000 ppm hydrogen
1000 ppm butane
1000 ppm propane
500 ppm n-octane
100 ppm hydrogen sulfide
15 ppm nitrogen dioxide
10 ppm sulfur dioxide
Hose and probe measurement
The following flushing time recommendations for hose and probe measurements are marked with a brand-new, dry and clean hose and probes
Hose Length (meter) Flashing time (min)
FKM Hose 5-20 3
FKM hose 45 5
PTFE-lined PVC hose 3-15 3
Bar probe 90 (83 16 532) 2
Telescopic probe ES 150 (83 16 533) 2
Probe Set GP600 (83 28 667) 3