Formaldehyde, 5 - 150 ppb (8610540)
Formaldehyde, 5 - 150 ppb, Part # 8610540
Measuring Range:
5 to 150 ppb
Typical Measuring Time:
approx. 360 to 960 s
Accuracy:
± 25 % (under calibration conditions)
Ambient Conditions:
Temperature:
15 to 35 °C
Temperature correction:
none
Humidity:
4 to 18 mg/L
Humidity correction:
< 8 mg/L 40 % from measurement value per mg/L
8 to 12 mg/L none
> 12 mg/L -10 % from measurement value per mg/L
Cross Sensitivity:
No influence at 75 ppb formaldehyde by: 0.5 ppm acrolein 10 ppm vinyl acetate 100 ppm styrene 100 ppb acetaldehyde
Hose and probe measurement
The following flushing time recommendations for hose and probe measurements are marked with a brand-new, dry and clean hose and probes
Hose Length (meter) Flashing time (min)
FKM Hose 5-20 3
FKM hose 45 5
PTFE-lined PVC hose 3 3
PTFE-lined PVC hose 15 10
Bar probe 90 (83 16 532) 1
Telescopic probe ES 150 (83 16 533) 2
Probe Set GP600 (83 28 667) 3