Horiba LAQUAtwin Pocket Meters
Horiba LAQUAtwin Pocket Meters
Your lab-in-a-pocket. Available for pH, Conductivity, Salt, Sodium-Ion, Potassium-Ion, Nitrate-Ion or Calcium-Ion.
HORIBA's unique compact meter integrates the electrode, display and sample container to enable simple, effective on-site testing by direct measurement from a single drop.
Unique sensor allows measurement of viscous liquids, solids, and even powder samples. No need for a beaker to calibrate your meter or measure a sample. Just place few drops of the standard or sample onto the flat sensor. This procedure saves you time and prevents wasting your precious sample.
- Auto calibration
- Auto temperature compensation
- Auto stable / auto hold measurement
- Auto power off: 30 mins
- Low battery indicator
- IP67 Rating dust & waterproof
- Replaceable sensor
- 400 Hours battery life: continuous use without backlight
- Material: ABS epoxy
- Display: custom (monochrome) LCD with backlight
- Operating conditions: 5 to 40°C, 85% or less relative humidity
- Warranty: 2 years - meter; 6 months - sensor
- Packed in handy carrying case
Models & Specifications
pH-11
- Glass Electrode
- 2 Calibration Points
- Minimum sample volume: 0.1ml (0.05ml with sampling sheet B)
- pH Range / Resolution: 0.0 to 14.0 pH / 0.1 pH
- Accuracy: +/- 0.1 pH
- pH Calibration Curves: USA - 1.68, 4.01, 7.00, 10.01 & 12.46. NIST - 1.68, 4.01, 6.86, 9.18 & 12.46
- mV Range / Resolution: +/- 650 mV / 1 mV
- No Temperature Display
Includes: 4.01 & 7.00 pH buffers (14 mL each), batteries, dropper, manuals & case
pH-22
- Glass Electrode
- 3 Calibration Points
- Minimum sample volume: 0.1ml (0.05ml with sampling sheet B)
- pH Range / Resolution: 0.00 to 14.00 pH / 0.01 pH
- Accuracy: +/- 0.01 pH
- pH Calibration Curves: USA - 1.68, 4.01, 7.00, 10.01 & 12.46. NIST - 1.68, 4.01, 6.86, 9.18 & 12.46
- mV Range / Resolution: +/- 650 mV / 1 mV
- No Temperature Display
Includes: 4.01 & 7.00 pH buffers (14 mL each), batteries, dropper, manuals & case
pH-33
- Glass Electrode
- 5 Calibration Points
- Minimum sample volume: 0.1ml (0.05ml with sampling sheet B)
- pH Range / Resolution: 0.00 to 14.00 pH / 0.01 pH
- Accuracy: +/- 0.01 pH
- pH Calibration Curves: USA - 1.68, 4.01, 7.00, 10.01 & 12.46. NIST - 1.68, 4.01, 6.86, 9.18 & 12.46
- mV Range / Resolution: +/- 650 mV / 1 mV
- Temperature Display / Resolution: 0 to 50.0° C / 0.1° C
Includes: 4.01 & 7.00 pH buffers (14 mL each), batteries, dropper, manuals & case
EC-11 (Conductivity)
- 2 Electrode Bipolar AC
- 2 Calibration Points
- Minimum sample volume: 0.12mL
- Conductivity Range / Resolution: 0 to 199 uS/cm / 1 uS/cm, 200 to 1999 uS/cm / 1 uS/cm, 2.00 to 19.99 mS/cm / 0.01 mS/cm
- Accuracy: +/- 2% Full Scale (for each range)
- Calibration Curves: 1413 uS/cm, 12.88 mS/cm
- No Temperature Display
Includes: 1413 uS/cm & 12.88 uS/cm Conductivity Standard Solutions (14 mL each), conditioning solution (4 mL), batteries, dropper, manuals & case
EC-22 (Conductivity)
- 2 Electrode Bipolar AC
- 3 Calibration Points
- Minimum sample volume: 0.12mL
- Conductivity Range / Resolution: 0 to 199 uS/cm / 1 uS/cm, 200 to 1999 uS/cm / 1 uS/cm, 2.00 to 19.99 mS/cm / 0.01 mS/cm, 20.0 to 199.9 mS/cm / 0.1 mS/cm
- Accuracy: +/- 2% Full Scale (0 to 19.99 mS/cm), +/- 5% FS (20.0 to 199.9 mS/cm)
- Calibration Curves: 1413 uS/cm, 12.88 mS/cm, 111.8 mS/cm
- Temperature Display / Resolution: 0 to 50.0° C / 0.1° C
Includes: 1413 uS/cm & 12.88 uS/cm Conductivity Standard Solutions (14 mL each), conditioning solution (4 mL), batteries, dropper, manuals & case
EC-33 (Conductivity)
- 2 Electrode Bipolar AC
- 3 Calibration Points
- Minimum sample volume: 0.12mL
- Conductivity Range / Resolution: 0 to 199 uS/cm / 1 uS/cm, 200 to 1999 uS/cm / 1 uS/cm, 2.00 to 19.99 mS/cm / 0.01 mS/cm, 20.0 to 199.9 mS/cm / 0.1 mS/cm
- TDS Range / Resolution: 0.0 to 99.9 ppm / 0.1 ppm, 100 to 999 ppm / 1 ppm, 1000 to 9990 ppm / 10 ppm
- Accuracy: +/- 2% Full Scale (0 to 19.99 mS/cm), +/- 5% FS (20.0 to 199.9 mS/cm)
- Calibration Curves: Conductiviity - 1413 uS/cm, 12.88 mS/cm, 111.8 mS/cm, TDS - TDS Factor (0.4 to 1.0) / EN 27888 / 442 / NaCL
- Temperature Display / Resolution: 0 to 50.0° C / 0.1° C
Includes: 1413 uS/cm & 12.88 uS/cm Conductivity Standard Solutions (14 mL each), conditioning solution (4 mL), batteries, dropper, manuals & case
Salt-11
- 2 Electrode Bipolar AC
- 2 Calibration Points
- Minimum sample volume: 0.12mL
- Salt Range / Resolution: 0.0 to 100.0 g/L / 0.1 g/L, 0.00 to 10.00% / 0.1%
- Accuracy: +/- 2% Full Scale (0.0 to 9.9 g/L), +/- 4% FS (10.0 to 100.0 g/L)
- Calibration Curves: NaCL / Sea water
- Temperature Display / Resolution: 0 to 50.0° C / 0.1° C
Includes: 0.5% & 5.0% NaCL Standard Solutions (14 mL each), conditioning solution (4 mL), batteries, dropper, manuals & case
Salt-22
- Ion Selective Electrode
- 2 Calibration Points
- Minimum sample volume: 0.3mL (0.05mL with sampling sheet B)
- Salinity Range: 0.01 to 25% by weight
- Resolution: 0.00 to 0.99% / 0.01%, 1.0 to 9.9% / 0.1%, 10 to 25% / 1%
- Accuracy: +/- 10% or reading value
- Temperature Display / Resolution: 0 to 50.0° C / 0.1° C
Includes: 0.5% & 5.0% NaCL Standard Solutions (14 mL each), conditioning solution (4 mL), batteries, dropper, manuals & case
Na-11 (Sodium Ion)
- Ion Selective Electrode
- 2 Calibration Points
- Minimum sample volume: 0.3mL (0.05mL with sampling sheet B)
- Range: 2 to 9900 ppm (mg/L) (0.1 to 430 mmol/L)
- Resolution: 0 to 99 ppm - 1 ppm, 100 to 990 ppm - 10 ppm, 1000 to 9900 ppm - 100 ppm
- Accuracy: +/- 10% of actual value
- Temperature Display / Resolution: 0 to 50.0° C / 0.1° C
Includes: 150 & 2000 ppm Standard Solution (14 mL each), Sampling Sheet B (5 pieces), batteries, dropper, manuals & case
K-11 (Potassium Ion)
- Ion Selective Electrode
- 2 Calibration Points
- Minimum sample volume: 0.3mL (0.05mL with sampling sheet B)
- Range: 4 to 9900 ppm (mg/L) (0.1 to 430 mmol/L), 2 to 5000 kg/10a (soil/water ratio 1:5)
- Resolution: 0 to 99 ppm - 1 ppm, 100 to 990 ppm - 10 ppm, 1000 to 9900 ppm - 100 ppm
- Accuracy: +/- 10% of actual value
- Temperature Display / Resolution: 0 to 50.0° C / 0.1° C
Includes: 150 & 2000 ppm Standard Solution (14 mL each), Sampling Sheet B (5 pieces), batteries, dropper, manuals & case
NO3-11 (Nitrate Ion)
- Ion Selective Electrode
- 2 Calibration Points
- Minimum sample volume: 0.3mL (0.05mL with sampling sheet B)
- Range: NO3 - 6 to 9900 ppm (mg/L) (0.1 to 160 mmol/L), NO3-N - 1.4 to 2200 ppm (mg/L)
- Resolution: 0 to 99 ppm - 1 ppm, 100 to 990 ppm - 10 ppm, 1000 to 9900 ppm - 100 ppm
- Accuracy: +/- 10% of actual value
- Temperature Display / Resolution: 0 to 50.0° C / 0.1° C
Includes: 150 & 2000 ppm Standard Solution (14 mL each), Sampling Sheet B (5 pieces), batteries, dropper, manuals & case
Ca-11 (Calcium Ion)
- Ion Selective Electrode
- 2 Calibration Points
- Minimum sample volume: 0.3mL (0.05mL with sampling sheet B)
- Range: 4 to 9900 ppm (mg/L) (0.1 to 250 mmol/L)
- Resolution: 0 to 99 ppm - 1 ppm, 100 to 990 ppm - 10 ppm, 1000 to 9900 ppm - 100 ppm
- Accuracy: +/- 20% of actual value
- Temperature Display / Resolution: 0 to 50.0° C / 0.1° C
Includes: 150 & 2000 ppm Standard Solution (14 mL each), Sampling Sheet B (5 pieces), batteries, dropper, manuals & case