Toluene, 10 - 1000 ppm (8610250)
Toluene, 10 - 1000 ppm Part # 8610250
Measuring Range:
10 to 1000 ppm
Typical Measuring Time:
6 to 360 s
Accuracy:
± 20 % (under calibration conditions)
Ambient Conditions:
Temperature:
0 to 40 °C
Temperature correction:
< 10 °C: +4% from the measured valve per °C
10 °C to 40 °C none
Humidity:
1 to 40 mg/L
Humidity correction:
none
Cross Sensitivity:
No influence at 50 ppm toluene by:
10 ppm xylene, isomer mixture
10 ppm p-xylene
25 ppm o-xylene
25 ppm benzene
300 ppm n-octane
Hydrogen Sulphide leads to considerable plus errors, therefore mercury measurements in the presence of hydrogen Sulphide are not possible.
Hose and probe measurement
The following flushing time recommendations for hose and probe measurements are marked with a brand-new, dry and clean hose and probes.
Hose Length (meter) Flashing time (min)
FKM Hose 5-20 3
FKM hose 45 7
PTFE-lined PVC hose 3-15 3
Bar probe 90 (83 16 532) 2
Telescopic probe ES 150 (83 16 533) 3
Probe Set GP600 (83 28 667) 3